Operating features, Input and output jacks – Martel Electronics MasterCAL User Manual

Page 22

Advertising
background image

MasterCAL

Users Manual

12

Operating Features

Input and Output Jacks

Figure 5 shows the calibrator input and output jacks.
Table 2 explains their use.

Table 2. Input/Output Jacks and Connectors

No. Name

Description

1

Battery Eliminator jack

Jack for the Model BE9005 Battery Eliminator. Use the battery eliminator for bench-top
applications where ac line power is available. This input does not charge the battery.

2

w

SERIAL PORT

Connects the calibrator to an RS-232 serial port on a personal computer.

3 Pressure

module

connector

Connects the calibrator to a pressure module.

4

TC input/output

Jack for measuring or simulating thermocouples. This jack accepts a miniature
polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in) center to
center.

5, 6

w

MEAS V jacks

Input jacks for measuring voltage, frequency, or three- or four-wire RTDs (Resistance
Temperature Detectors).

7, 8

w

SOURCE mA,

MEAS mA

RTD

jacks

Jacks for sourcing or measuring current, measuring resistance and RTDs, and
supplying loop power.

9,10

w

SOURCE V

RTD

jacks

Output jacks for sourcing voltage, resistance, frequency, and for simulating RTDs.

Advertising