The procedure for testing a pressure limit switch
follows. The switch in this example sets at a high
limit of 10 psi. The set state is a closed switch
contact. For pressure switches, you use the

Manual

Test

choice. For testing switches that do not require

sourcing pressure, you can use the

Auto Test

choice.

1.

Connect the test leads between the pressure
switch contact output and the mA

RTD

(middle) jacks on the calibrator.

2.

Connect the pressure module to the calibrator,
and connect a pressure line to the limit switch.
Leave the pressure line vented to atmosphere.

3.

If necessary, press

M

for MEASURE mode.

4.

Press

q q

for the continuity measure

function.

5.

Press

M

for SOURCE mode.

6.

Press

p

for the pressure source function.

7.

Press

c

to zero the pressure module.

8.

Press

M

.

9.

Press the

As Found

softkey.

10.

Highlight

1 Pt. Switch Test

from the menu and

press

e

.

11.

Press

e

to modify the parameters for

Setpoint 1.

12.

Make the following selections:

Setpoint 1 = 10.000 psi
Setpoint Type = High
Set State = Short

13.

Press the

Done

softkey.

14.

Set the

Tolerance

to 0.5 psi.

15.

The next parameters,

Deadband Min

and

Deadband Max

, are optional. Do not set them in

this example.