FIas leftfl test run, As left” test run – Martel Electronics MasterCAL User Manual

Page 87

Advertising
background image

Calibrator

Calibrating a Process Instrument

77

4.

Adjust the transmitter output for 20 mA.

5.

If the span was adjusted in step 4, you must go
back and repeat steps 3 and 4 until no more
adjustment is required.

6.

Now check the transmitter at 50%. If it is within
specification, your adjustment is complete. If not,
adjust the linearity and begin this procedure
again at step 3.

“As Left” Test Run

Proceed as follows to generate and record

as left

data for the thermocouple temperature transmitter
you have just adjusted.

1.

Press the

As Left

softkey to record

as left data.

2.

Press the

Auto Test

softkey to begin an

automatic sequence through all the test points,
or you can step through the tests manually.

3.

When the tests are complete, observe the error
summary table, such as the following.

gj48s.eps

An asterisk (

*

) next to a measure or source value

indicates an unsettled value (

g

annunciator)

when the measurement was taken.

4.

If all the results are within specification, as they
are this time, press the

Done

softkey. An entry in

memory is made for

as left data.