Martel Electronics MasterCAL User Manual

Page 42

Advertising
background image

MasterCAL

Users Manual

32

1.

Connect a pressure module to the calibrator as
shown in Figure 11. The threads on the
pressure modules accept standard ¼ NPT pipe
fittings. Use the supplied ¼ NPT to ¼ ISO
adapter if necessary.

2.

Press

M

for MEASURE mode.

3.

Press

p

. The calibrator automatically senses

which pressure module is attached and sets its
range accordingly.

4.

Zero the pressure module as described in the
module’s Instruction Sheet. Modules vary in
zeroing procedures depending on module type.
You MUST perform this step before you execute
a task that sources or measures pressure.

5.

If desired, you can change pressure display
units to psi, mHg, inHg, mH

2

O, inH

2

O@,

inH

2

O@60

°

F, ftH

2

O, bar, g/cm

2

, or Pa. Metric

units (kPa, mmHg, etc.) are shown in Setup
mode in their base units (Pa, mHg, etc.).
Change the pressure display units as follows:

a.

Press

s

.

b.

Press

Next Page

twice.

c.

Press

e

or the

Choices

softkey with

the cursor on

Pressure Units

.

d.

Select the pressure units with

u

or

d

.

e.

Press

e

.

f.

Press

Done

.