MasterCAL

Users Manual

46

Using Source Mode

The operating mode (i.e., MEASURE, SOURCE) is
shown in a reverse-video bar on the display. If the
calibrator is not in SOURCE mode, press

M

until

SOURCE is shown. You must be in SOURCE mode
to change any of the SOURCE parameters.

Sourcing Electrical Parameters

To select an electrical sourcing function, proceed as
follows:

1.

Connect the test leads as shown in Figure 16,
depending on the source function.

2.

Press

m

for current,

v

for dc voltage,

h

for frequency, or

q

for resistance.

3.

Enter the desired output value, then press

e

. For example, to source 5.0 V dc, press

v

5

.

0

e

.

Note

If you are sourcing frequency, respond to the
display prompt to select a zero-symmetric
sine or positive square wave. The amplitude
you specify is p-p amplitude.

4.

To change the output value enter a new value
and press

e

.

Note

If you are sourcing current, wait for the

g

symbol to go out before you use the output.

5.

To set the output value to 0 in the present
source function, press

c

.

6.

To turn off sourcing completely, press

c

twice.

Note

Use the source current function to drive a
current loop. This is different than the loop
power function in which the calibrator is
powering a process instrument. To source
loop power, use the

Loop Power

function

accessible from Setup mode.