MasterCAL

Users Manual

78

Test Comments

The calibrator runs tasks (custom procedures) that
are developed using a host computer and compatible
application software. A task may display a list of
proposed comments during execution. When the
comment list is displayed, select a comment to be
saved with the test results by pressing the

u

and

d

keys followed by

e

.

Calibrating a Delta-Pressure Flow
Instrument

The procedure to calibrate a

instrument is the

same as for other instruments, as just described,
with the following differences:

Source square-root is automatically enabled
after the

As Found

calibration template is

complete.

Measure/Source displays are in engineering
units.

The measurement percentage is automatically
corrected for the transmitter’s square-root
response, and is used to compute instrument
errors.

You select the

instrument procedure in a menu

after you press the

As Found

softkey.