Measuring or sourcing with custom units – Martel Electronics MasterCAL User Manual

Page 53

Advertising
background image

Calibrator

Using Measure Mode

43

Measuring or Sourcing with Custom
Units

w

Warning

To avoid possible electric shock, when
using Custom Units for measurement,
always refer to the secondary value
displayed below and to the right of the
main display for the actual value of the
measurement in native engineering units.

You can set up the measurement or source display
to show your own custom units. To do this, you
select a function, for example mV dc, scale it as you
choose, then enter an alphanumeric name for your
custom units, for example, “PH.”

Proceed as follows to set up a custom unit:

1.

When measuring or sourcing the function of your
choice, press the

Scale

softkey, then select

Custom Units

from the list.

2.

Enter the 0% and 100% scale points for the input
of your transfer function.

3.

Press the

Custom Units

softkey.

4.

Enter the 0% and 100% scale points for the
output of your transfer function.

5.

Enter the name of the custom units (up to four
characters), for example

PH

(for pH), using the

alphanumeric entry window, then press

e

.

While

Custom Units

are active, the

w

symbol shows

on the display to the right of the custom unit. Once
you have programmed the custom measurement
unit, the custom unit is available for calibration
procedures in split-screen MEASURE/SOURCE
mode. To cancel

Custom Units

, press the

Custom Units

softkey again.