Adjusting the transmitter – Martel Electronics MasterCAL User Manual

Page 86

Advertising
background image

MasterCAL

Users Manual

76

19.

In the results summary test, failures are
highlighted. An adjustment is required in this
example because three tests show failures. The
failures were outside the ±0.5% tolerance that
we selected.

20.

Either press the

Done

softkey to save the data,

or the

Abort

softkey to delete the data and start

over.

You can see the saved data entry and recall the
table for later viewing through the

Review

Memory

softkey during normal operation. You

can upload this data to a host computer running
compatible application software.

Adjusting the Transmitter

Proceed as follows to make the calibration
adjustments to the transmitter. (Always refer to the
transmitter manufacturer’s instructions to locate the
adjustment controls and connection points for your
transmitter.)

1.

Press the

Done

softkey while viewing the results

summary.

2.

Press the

Adjust

softkey. The calibrator sources

0% of span (100°C in this example) and displays
the following softkeys:

Go to 100%/Go to 0%

Go to 50%

As Left

Exit Cal

3.

Adjust the transmitter output for 4 mA then press
the Go to 100% softkey.

Advertising