5 firmware upgrade instructions, 5 firmware upgrade instructions -2, Boonton 4240 series rf power meter – Boonton 4240 RF Power Meter User Manual

Page 122: Maintenance 7-2

Advertising
background image

Boonton 4240 Series RF Power Meter

7.5 Firmware Upgrade Instructions

equirements

The Firmware Update file is a text file in S-record format. Any utility that can send a text file over
the GPIB interface could be used, however, it is recommended that National Instruments VISA
Interactive Controller (VIC) utility that is available with installation of their VISA (Virtual
Instrument Software Architecture) product be used as this has been tried and tested. Please refer to
the specific vendor’s website for license and download details.

rocedure

1. Save the file "Srecord_4240.hex" to your hard drive in a folder such as "C:\4240\Update"

2. On the 4240 configure the following IEEE parameters as shown;

EMULATION

4230

EOS

LSTN

LF

3. Start the VISA Interactive Controller utility


R

P

4. Double click on the 4240 Resource (in this example ‘GPIB0::13::INSTR’) to open a VISA

Session. Ensure that the ‘Show All VISA Operations’ box is checked. Then click the ‘Basic
I/O’ tab and then the ‘viWriteFromFile’ tab on that form.

5. Set the count to 2000000 then either Browse for or enter the Filename.

6. Click the ‘Execute’ button to send the file to the instrument.

Maintenance

7-2

Advertising