Sense:filter:state, Sense:filter:time, Sense:filter:state -25 – Boonton 4240 RF Power Meter User Manual

Page 85: Sense:filter:time -25, Boonton 4240 series rf power meter, Remote operation 5-25

Advertising
background image

Boonton 4240 Series RF Power Meter

SENSe:FILTer:STATe

Description:

Set or return the current setting of the integration filter on the selected channel. OFF provides no
filtering, and can be used at high signal levels when absolute minimum settling time is required.
ON allows a user-specified integration time, from 50 milliseconds to 20 seconds (see
SENSe:FILTer:TIMe

command). Note that setting the filter time will force the state to ON.

t of filtering, which is set automatically by the power meter based on

that gives a good compromise between measurement noise and

SENSe[1|2]:FILTer:STATe <character data>

Argument:

<character data> = {OFF, ON, AUTO}

SENSe:FILTer:TIMe

Description:

rn the current length of the integration filter on the selected channel. If the filter state is

d if set to OFF, a time query will return 0.00.
ON.

meric value>

econds in 50 millisecond increments

AUTO uses a variable amoun
the current signal level to a value
settling time at most levels.

Syntax:

Set or retu

set to AUTO, querying the time will return -0.01, an
Note that setting the filter time will force the state to

Syntax: SENSe[1|2]:FILTer:TIMe <nu

Argument:

<numeric value> = 0.05 to 20.00 s

Remote Operation

5-25