5 fetch queries, Fetch:cw:power, Fetch:key – Boonton 4240 RF Power Meter User Manual

Page 77: 5 fetch queries -17, Fetch:cw:power? -17, Fetch:key? -17, Boonton 4240 series rf power meter, Remote operation 5-17

Advertising
background image

Boonton 4240 Series RF Power Meter

5.5.5 FETCh Queries

The FETCh? group of queries is used to return specific measurement data from a measurement cycle that has been INITiated
and is complete or free-running. FETCh? performs the data output portion of the measurement. FETCh? does not start a
new measurement, so a series of FETCh? queries may be used to return more than one set of processed measurements from a
complete set of acquired data. FETCh? usually returns the current value of measurements, and should be used anytime free
running data acquisition is taking place (INITiate:CONTinuous ON). If FETCh? is used for single measurements
(INITiate:CONTinuous OFF), no data will be returned until a measurement has been INITiated and is complete.


FETCh:CW:POWer?

Syntax:

asurement condition code.

ETCh:KEY?

Return the key code of the last key depressed; e.g. MENU = 1.

Returns:

de

FREQ

4

AVG

8

Zero/Cal

16

REF

Level

32

Up

Arrow

64

Left

Arrow

128

Enter

256

Right

Arrow

512

Down

Arrow

1024







Description:

Return current average amplitude reading in channel units.

FETCh[1|2]:CW:POWer?

Returns:

CC, average power

Where CC is the me


F

Description:

Syntax:

FETCh:KEY?

key

code

Key

Co

Menu

1

Sensor

2

Remote Operation

5-17