5.5.7 MEASure Queries

used to acquire data using a set of high level instructions. They are structured to allow

e measurement process for easy operability. MEASure? provides a complete capability

ecific format.

tion. The instrument

Returns:

CC, Power in dBm

Where CC is the measurement condition code.

?

efault instrument configuration in volts units. Instrument remains

Syntax: MEASure[1|2]:VOLTage?

Returns:

CC, Average voltage in linear volts

CC is the measurement condition code.

The MEASure group of commands is
the user to trade off fine control of th
where the power meter is configured, a measurement taken, and results returned in one operation. The instrument is set to a
basic, predefined measurement state with little user intervention necessary or possible. Sometimes, more precise control of
measurement is required. In these cases, MEASure? should not be used. Rather, a sequence of configuration commands,
generally from the CALCulate and SENSe groups should be used to set up the instrument for the measurement, then READ?
or FETCH? commands are used to return the desired measurement data in a sp

MEASure:POWer?

Description:

Return a power measurement in dBm using a default instrument configura
remains stopped after a measurement.

Syntax:

Measure[1|2]:POWer?

MEASure:VOLTage

Description:

Return average voltage using a d
stopped after a measurement.

Where