Boonton 4240 Series RF Power Meter
CALCulate:MATH:ARGA
Description:
Set or return the first argument to be used for channel math operations.
Syntax:
CALCulate:MATH:ARGA <character data>
Argument:
<character data> = { CH1, CH2 }
CALCulate:MATH:ARGB
Syntax:
H:ARGB <character data>
:
?
Syntax:
ALCu
H:OPE
Description:
Set or return the channel math operator.
Syntax:
CALCulate:MATH:OPERator <character data>
Argument:
<character data> = { CH_SUM, CH_DIFF, CH_RAT }
ALCu
e
n:
ese measurement modes are only available
under bus control. NORMal is the default mode of operation at approximately 20 readings per
second. FAST returns data at over 200 readings per second. FILTered will only return data when
the filter is full. FILTered works best when INITiate:CONTinuous is OFF and a trigger is issued.
In this case the measurement will not be returned until the filter is full. When the instrument is
returned to local the MODE defaults to NORMal.
Syntax:
CALCulate:MODe <character data>
Argument:
<character data> = { NORMal, FAST, FILTered }
Description:
Set or return the second argument to be used for channel math operations.
CALCulate:MAT
Argument
<character data> = { CH1, CH2 }
CALCulate:MATH:DATA
Description:
Returns the channel math result.
CALCulate:MATH:DATA?
C
late:MAT
Rator
C
late:MOD
Descriptio
Set or return the system remote measurement mode. Th