5 scpi command reference, 1 ieee 488.2 commands, 5 scpi command reference -6 – Boonton 4240 RF Power Meter User Manual

Page 66: 1 ieee 488.2 commands -6, Cls -6, Ese -6, Boonton 4240 series rf power meter, Remote operation 5-6

Advertising
background image

Boonton 4240 Series RF Power Meter

5.5 SCPI Command Reference

te commands accepted by the 4240 Series. The list is grouped by SCPI subsystem

r IEEE

.5.1 IEEE 488.2 Commands

he purpose of IEEE488.2 commands is to provide management and data communication instructions for the system by

g a set of “*” commands (an asterisk followed by a three character code). These commands allow device control and

onitoring, and are e basis for some of the commands of the SCPI STATus subsystem (see Section 5.5.12).

*CLS

Description:

surement.


This section contains a list of all SCPI remo
o

488.2 function, and includes a detailed description of each command.

5

T
definin
status m

th

Clear Status command. This command resets the SCPI status registers (Questionable Status and
Operation Status), the error queue, the IEEE488.2 Status Byte (STB) and Standard Event Status
(ESR) registers, and the mea

Syntax:

*CLS

Argument:

None

*ESE

Description:

Set or return the Standard Event Status Enable Register. The mask value in this register is used to
enable bits of the Standard Event Status Register that are or’ed together to form the ESB summary
bit in the instrument Status Byte. When a mask bit is set, and the corresponding ESR bit goes
true, an SRQ will be generated, provided the Event Status Summary bit (ESB, bit 5) is enabled in
the SRE register. No SRQ can be generated for that condition if the mask bit is cleared. To clear
the entire Standard Event Status Enable Register, send *ESE = 0. See the *ESR command for bit
assignments. This register is not cleared by *CLS, *RST or DCLR.

Syntax:

*ESE <numeric value>

Argument:

<numeric value> = 0 to 255

Remote Operation

5-6

Advertising