USER'S GUIDE____________________________________________________________________

32 _______________________________________________________________ M210912EN-C

The settings are temporary, and last only for a single session at a time.
On the next connection HMP155 reverts to the settings last set with the
SERI command.

SMODE

Use the SMODE command to set the user port to permanent STOP,
RUN, POLL or SEND mode.

SMODE [xxxx]

where

xxxx= STOP, RUN, POLL or SEND

Example:

>smode
Serial mode : STOP ? POLL
>

Table 12

Selection of Output Modes

Mode Output

Available

Commands

STOP

Only with the SEND command

All (default mode)

RUN

Automatic output

Only command S

POLL

Only with the SEND [addr] command

??, SEND [addr], OPEN [addr]

Selected output mode will be activated after power outages.

INTV

Use the INTV command to set the outputting interval for the RUN mode.
The default value for this command is 2 seconds.

INTV [xxx yyy]

where

xxx =

Output interval (0...255) where 0 produces an output whenever a
new measurement is made. Interval 1 s produces repeats of the
previous measurement if no newer measurement is available.

yyy =

Unit (s, min or h)

Example:

>intv 1 min
Interval : 1 min
>