Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 37

System Commands

The commands introduced in this section are used for setting the filtering
level, checking the probe configuration, listing available commands and
possible error messages, and displaying the software information of the
probe.

FILT [0.1…1]

Use the FILT [xxx] serial line command to set the filtering level. The
default setting is 1. The output is calculated based on the following
formula: [( new result * filt) + ( old result * (1.0 - filt))]. This means that
with FILT value 1 the probe will take only the latest measurement into
account, but with FILT value 0.1 a new output is a combination of the
previous output (90%) and the latest measurement (10%).

FILT

[xxx]

where

xxx =

0.1…1 where 1.0 = no filter and 0.1 = about 16 moving average

Example:

>filt
Filter : 0.800 ?
>