Chapter 7 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 71

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the product.

Performance

Relative Humidity

Table 16

Relative Humidity Performance

Description Value
Measurement range

0 ... 100 %RH

Accuracy (including non-linearity, hysteresis and repeatability)

±1% RH (0 ... 90% RH)

at +15 ... 25 °C (59 … 77 °F)

±1.7% RH (90 ... 100% RH)

at -20 … +40 °C (-4 … 104 °F)

±(1.0 + 0.008 × reading)% RH

at -40 ... -20 °C (-40 … -4 °F)

±(1.2 + 0.012 × reading)% RH

at +40 … +60 °C (104 … 140 °F)

±(1.2 + 0.012 × reading)% RH

at -60 ... -40 °C (-76 … -40 °F)

±(1.4 + 0.032 × reading)% RH
±0.6% RH (0 ... 40% RH)

Factory calibration uncertainty
(+20 °C)

1

±1.0% RH (40 ... 97% RH)

Recommended humidity sensor
HUMICAP

®

180R

for typical applications

HUMICAP

®

180RC

for applications with chemical purge
and/or warmed probe

Response time for HUMICAP

®

180R(C) at 20 °C in still air with

sintered PTFE filter
63% 20

s

90% 60

s

1

Defined as ± 2 standard deviation limits. Small variations possible, see also calibration

certificate.