Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 61

One-Point Humidity and Temperature
Adjustment

To make a one-point humidity or temperature adjustment:

1.

Carry out the chemical purge (if available). See section Chemical
Purge (Optional) on page 45.

2.

Uncover the adjustment buttons on the probe by removing the
protective cover and removing the calibration seal.

3.

Open the protective plug and you can see three buttons marked -,
and +. There is also a two-colored indicator LED. See Figure 13 on
page 58.

4.

Remove the filter and insert the probe into a measurement hole of
the salt chamber to do the humidity adjustment.

5. Press

the

ADJ

button and hold it down until the green indicator

LED lights up. This puts HMP155 in RH calibration mode.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized.

6. Using

the

-

and + buttons, make sure the A

out

voltage is correct and

press the ADJ button. Green indicator LED turns off and back on.

7.

To move on to the temperature adjustment, press the ADJ button
once and the red indicator LED lights up.

8.

Insert the probe into the reference temperature.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized.

9. Using

the

-

and + buttons, make the temperature offset adjustment

by making sure the A

out

voltage is correct and press the

ADJ

button. Red indicator LED turns off and back on.

10. Press

the

ADJ

button one more time so that the red LED turns off

to indicate the probe has quit the calibration mode.