Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 65

Relative Humidity Adjustment after
Sensor Change

After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.

FCRH

Example

:


>fcrh
RH : 25.19 Ref1 ? 11.3
Press any key when ready ...
RH : 70.02 Ref2 ? 75.5
OK
>

The

OK

indicates that the calibration has been successful.

One-Point Humidity and Temperature
Adjustment

One-point humidity and temperature adjustment can be done by using the
following serial line commands:

- CRH and FCRH commands for humidity adjustment

- CT and CTA commands for temperature adjustment

To make an adjustment, just enter the command , press space bar and
then enter the reference value, as in the example below.

Example

:

Crh 20.0
OK