Atest, Aerr – Vaisala HMP155 User Manual

Page 45

Advertising
background image

Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 43

ATEST

Use the serial line to test the operation of all the analog outputs. Use the
ATEST

command to force the analog outputs to entered values. The set

values remain valid until you enter the ATEST command without
parameters or reset the probe. Entering the ATEST command without
parameters outputs the current output value.

ATEST [V] [V]

where

V

= Output value for channel 1 (V)

V

= Output value for channel 2 (V)

Example 1:

> atest
2.412 6301
1.943 5090
>

Example 2:

> atest 1.5 1.5
1.500 3948
1.500 3948
>

AERR

Use the AERR serial line command to set or view the error output from
analog outputs.

AERR [CH0ERR/CH1ERR]

where

CH0ERR

=

Analog

output

CH1ERR = Analog

output

Example:

>aerr
Ch0 error out : 10.000V ? 0
Ch1 error out : 1.000V ? 0
>

NOTE

The error output value must be within the valid range of the output
mode.

Advertising