Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 49

XHEAT

Enables or disables the sensor heating.

Example:

>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>

To configure the sensor heating, use the XHEAT command without
parameters. Enter the values after the question mark. The available
ranges include the following:

Extra heat RH -limit (heating
function starts-up above the
setpoint)

0 ... 100 %RH (default: 95 %RH)

Extra heating temperature

0 ... 200 ºC (default: 100 ºC)

Extra heating time

0 ... 255 s (default: 30 s)

To see the current values, use the XHEAT * command.

Example

:


>xheat *
Extra heat : OFF
RH limit : 95 %RH
Temperature : 100 'C
Duration : 30 s
>

NOTE

After the defined XHEAT duration, there is an additional cooling time of
10 seconds before outputs are updated.