Serial line calibration, Two-point humidity adjustment – Vaisala HMP155 User Manual

Page 65

Advertising
background image

Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 63

Serial Line Calibration

Two-Point Humidity Adjustment

Note that the difference between the two humidity references must be at
least 30 %RH.

To make a humidity adjustment:

1.

Connect the HMP155 to a PC. See section Serial Line
Communication on page 24. Open a term
inal program.

2.

Carry out the chemical purge (if available). See section Chemical
Purge (Optional) on page 45.

3. Press

the

ADJ

button and the green LED lights up.

4.

Remove the filter from the probe and insert the probe into a
measurement hole of the dry end reference chamber (for example,
LiCl: 11 % RH).

5. Enter

the

CRH

command and press ENTER.

6.

Wait at least 30 minutes for the sensor to stabilize.

7. Press

ENTER

a few times to check if the reading has stabilized.

8.

When the reading has stabilized, give the reference humidity after
the question mark and press ENTER.

>crh
RH : 16.6675 1. ref ?
RH : 16.4978 1. ref ?
RH : 16.3956 1. ref ? 11.25
Press any key when ready ...
RH : 11.25 Ref2 ?

9.

Now the device is waiting for the high end reference. Insert the
probe into the measurement hole of the high end reference chamber
(for example, NaCl: 75 % RH chamber in the humidity calibrator
HMK15). Press any key when ready.

10. Let the probe stabilize for about 30 minutes. You can follow the

stabilization by pressing ENTER.

11. When stabilized, type the high end reference value after the

question mark and press ENTER.

RH : 75.45 Ref2 ?
RH : 75.57 Ref2 ?
RH : 75.55 Ref2 ?
RH : 75.59 Ref2 ? 75.5
OK

12. OK indicates that the adjustment has succeeded and the new

calibration coefficients are calculated and stored.