Chapter 7 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 75

Accuracy of Wet Bulb Temperature °C

Relative

humidity

Temp.

10

20 30 40 50 60 70 80 90 100

-40

0.20

0.20 0.20 0.20 0.20 0.20 0.20 0.20

-20

0.21

0.21 0.22 0.22 0.22 0.22 0.23 0.23

0

0.27

0.28 0.28 0.29 0.29 0.29 0.30 0.30 0.31 0.31

20

0.45

0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42

40

0.84

0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52

60

1.45

1.20 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.60

Accuracy of Dew point Measurement

Find the intersection of the dew point temperature curve and the dew
point difference reading (process temperature-dew point temperature) on
the x-axis and read the accuracy in dew point measurement on the y-axis.

0508-017

Figure 15

Accuracy of Dew point Measurement

Inputs and Outputs

Table 20

Outputs

Description Value
Voltage output

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Resistive Pt100 4-wire connection

RS-485
Minimum operating voltages
0…5 V output

12 V

0…10 V output, probe heating,
chemical purge or XHEAT

16 V