Ctext, Cdate, Analog output adjustment – Vaisala HMP155 User Manual

Page 69

Advertising
background image

Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 67

commands in ADJUST mode. To put HMP155 in adjustment mode, just
hold down the ADJ button until the green LED lights up.

CTEXT

Use the CTEXT command to enter text to the adjustment information
field.

Example:

>ctext
Cal. info : Helsinki / FIN ? Finland
>

CDATE

Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYYMMDD.

Example:

>cdate 20080320
>

Analog Output Adjustment

In the analog output calibration the analog output is forced to the
following values:

- Voltage output: 10 % and 90 % of the range

Connect HMP155 to a calibrated voltage meter in order to measure
voltage.

NOTE

Hold down the ADJ button until a green LED lights up. This puts the
HMP155 in ADJUST mode.

NOTE

The serial line command ACAL cannot be used with HMP155 passive
output version.

Enter the ACAL command and type the multimeter reading for each
case. Continue by pressing ENTER. When a channel is specified, only
the specified analog output channel is adjusted.

Advertising