APPENDIX A

CALCULATION FORMULAS

This Appendix contains the formulas used for the calculated output
quantities.

The HMP155 series probe measures relative humidity and temperature.
From these values dew point and frost point, mixing ratio, wet bulb
temperature, absolute humidity and enthalpy in normal pressure are
calculated using the following equations:

Dew point/frost point:

15

.

273

1

1078

.

6

log

5

.

7

3

.

237

10

/





w

f

d

P

T

(1)


P

w

is the water vapor pressure. Dew point/frost point is measured in

Kelvin.


Mixing ratio:

w

w

P

p

P

x

-

99

.

621

(2)

Absolute humidity:

T

P

C

A

w

(3)

where

C = 216.679