Vaisala HMP155 User Manual

Page 62

Advertising
background image

USER'S GUIDE____________________________________________________________________

60 _______________________________________________________________ M210912EN-C

7.

Insert the probe into the high end reference chamber (for example,
NaCl: 75 % RH chamber in the humidity calibrator HMK15) and
do the high humidity gain adjustment by using the - and + buttons
to make sure the A

out

voltage is correct. To finish the RH

calibration, press the ADJ button and the red indicator LED lights
up.

NOTE

If you do not wish to perform the temperature adjustment at this time,
press ADJ -button two times and the indicator LED turns off. After
reset, HMP155 exits the calibration mode. Otherwise, continue
following the instructions from step 8 onwards.

8.

Insert the probe into a known reference temperature (if HMK15
Humidity Calibrator is not used) and let the temperature reading
stabilize.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized.

9. Using

the

-

and + buttons, make the temperature offset adjustment

by making sure the A

out

voltage is correct and press the ADJ

button. Red indicator LED turns off and back.

NOTE

If you do not wish to perform the two-point temperature adjustment at
this time, press the ADJ button one more time so that the red indicator
LED turns off. After reset, HMP155 exits the calibration mode.
Otherwise, continue following the instructions from step 10 onwards.

10. Insert the probe into another reference temperature.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized.

11. Using

the

-

and + buttons, make the temperature gain adjustment

by making sure the A

out

voltage is correct.

12. Press

the

ADJ

button and the red indicator LED turns off.