Push-button calibration, Two-point humidity and temperature adjustment – Vaisala HMP155 User Manual

Page 61

Advertising
background image

Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 59

Push-Button Calibration

Calibration and adjustment of the HMP155 can be done by using the
adjustment buttons found on the probe. The two-point humidity
adjustment is carried out by using two relative humidity references: for
example saturated salt points 11 % RH (LiCl) and 75 % RH (NaCl). The
difference between the two humidity reference points must be at least 30
%RH. The difference between the two temperature reference points must
be at least 30 °C.

NOTE

In push-button calibration, the probe outputs are by default relative
humidity in the range 0…100 %RH in CH1 and temperature in the range
-40…+60°C in CH2.

Two-Point Humidity and Temperature
Adjustment

To make a humidity or temperature adjustment:

1.

Carry out the chemical purge (if available). See section Chemical
Purge (Optional) on page 45.

2.

Uncover the adjustment buttons on the probe by removing the
protective cover and removing the calibration seal.

3.

Open the protective plug and you can see three buttons marked -,
and +. There is also a two-colored indicator LED. See Figure 13 on
page 58.

4. Press

the

ADJ

button and hold it down until the green indicator

LED lights up. This puts HMP155 in RH calibration mode.

5.

Remove the filter and insert the probe into a measurement hole of
the dry end reference chamber (for example, LiCl: 11 % RH) to do
the low humidity offset adjustment.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized. This takes approximately 30 minutes.

6. Using

the

-

and + buttons, make sure the A

out

voltage is correct and

press the ADJ button. Green indicator LED turns off and back on.

NOTE

Do not touch the adjustment buttons before the conditions have
stabilized.