Temperature – Vaisala HMP155 User Manual

Page 74

Advertising
background image

USER'S GUIDE____________________________________________________________________

72 _______________________________________________________________ M210912EN-C

Temperature

Table 17

Temperature Performance

Description Value
Measurement range

-80 ... +60 °C (-112 ... +140 ºF)

Accuracy with voltage output
at -80 … +20 °C

±(0.226 - 0.0028 x temperature) °C

at +20 … +60 °C

±(0.055 + 0.0057 x temperature) °C

Accuracy with passive (resistive) output
According to IEC 751 1/3
Class B

±(0.1 + 0.00167 x |temperature|) °C

Accuracy with RS-485 output (see Figure 14 below)
at -80 … +20 °C

±(0.176 - 0.0028 x temperature) °C

at +20 … +60 °C

±(0.07 + 0.0025 x temperature) °C

Temperature sensor(s)

Pt 100 RTD Class F 0.1 IEC 60751

Response time (63%) for additional temperature probe in 3 m/s air flow
63%

< 20 s

90%

< 35 s

0804-032

Figure 14

Accuracy over Temperature Range: Voltage and RS-
485