Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 33

Formatting Commands

FORM

Use the serial line command FORM to change the format or select
certain quantities for the output commands SEND and R.

FORM [x]

where

x = Formatter

string

Formatter string consists of quantities and modifiers. You can only give a
maximum of 73 characters after the command on the command line.

When entering the command, use the abbreviations of the quantities. For
more information on quantities, see Table 3 and Table 4 on page 11.

The modifiers are presented in Table 13 below.

Table 13

FORM Command Modifiers

Modifier Description
quantity

Quantity name (for example RH, T or TDF)

x.y

Length modifier (number of digits and decimal places)

#t Tabulator
#r Carriage-return
#n Line

feed

“" String

constant

#xxx

Special character with decimal code "xxx"

U5

Unit field and length

ADDR

Probe address with two characters [00...99]

CS2

Modulus-256 checksum of message sent so far,
hexadecimal format.

CS4

Modulus-65536 checksum of message sent so far,
hexadecimal format.

ERR

Error flags for P, T, Ta, RH [0000 ... 1111], 0 = no error

STAT

Probe status in 7 character field, for example:
N no

heating

h

probe heating active, power

H

purge heating active, temperature

S

purge cooling active, temperature

X

sensor heating active, temperature

SNUM Probe

serial

number

TIME Time

[hh:mm:ss]

Example:

>form "Temperature=" 5.2 t #r#n