USER'S GUIDE____________________________________________________________________

66 _______________________________________________________________ M210912EN-C

User Adjustment Commands

You can use the user adjustment commands to display and set
parameters. You can only use LI command in ADJUST mode.

These parameters are updated with the commands CRH, CT, and CTA.

L

With the L command you can display user adjustment parameters.

Example:

>l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Ta offset : 0.00000000E+00
Ta gain : 1.00000000E+00
>

LI

With the LI command you can display and ask for new values for the
user adjustment parameters.

Example:

>li
Cp offset : 0.00000000E+00 ?
Cp gain : 1.00000000E+00 ?
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
Ta offset : 0.00000000E+00 ?
Ta gain : 1.00000000E+00 ?
>

NOTE

If you suspect that something has gone wrong with the calibration, you
can use the LI command to return the factory values by entering the
values shown in the example above.

Feeding Adjustment Information

The information added with the serial line commands below is shown in
the device information fields. You can only use CTEXT and CDATE