RS232

ASCII Interface

Protocol

jins85e1 2.fm

Interface Description

(1101)

Command

word1

Command

word2

Command

word3

Parameter

Statement

Read/

Set

Example

Response

*CONFig

:PLCOUTLINK

PLC output link for one pin

:<no.>

pin number = 8 .. 15

R/S

<value>

values for PLC output functions:

R/S

OPEN

S
R

*CONF:PLCOUTLINK:9 OPEN
*CONF:PLCOUTLINK:9?

OK

OPEN

CLOSE

TRIGGER_1 INV_TRIGGER_1
TRIGGER_2 INV_TRIGGER_2
READY INV_READY
CYCLE_ACTIVE INV_CYCLE_ACTIVE
STANDBY INV_STANDBY
ERROR INV_ERROR
WARNING INV_WARNING
ERR_WARN INV_ERR_WARN
CAL_ACTIV INV_CAL_ACTIV
MEASURE INV_MEASURE
GROSS INV_GROSS
CONTROL_OUT_FUNC_GR_LEAK
CONTROL_OUT_FUNC_INV_GR_LEAK

R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S
R/S

:PROOFleak

read/set proof leak rate

S

R

*CONF:PROOF 6E-6
*CONF:PROOF?

OK

6.00E-6

<value>

value = MBAR*L/S, PA*M3/S, SCCM, TORR*L/
S

S
R

*CONF:PROOF:MBAR*L/S 1.5E-5
*CONF:PROOF:MBAR*L/S?

OK

1.50E-5

Advertising