INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 16

Advertising
background image

16

RS232

ASCII Interface

Protocol

jins85e1 2.fm

Interface Description

(1101)

Command

word1

Command

word2

Command

word3

Parameter

Statement

Read /

Set

Example

Response

*STATus

read status of the unit

deprecated! Use STAT:MEAS instead!

R

INIT, START, MEAS, CAL, ERROR, ADJUST,

STANDBY, OVERRANGE

:BackGND

report helium background quality

Possible values = OK, Moderate, Bad

R

*STAT:BGND?

OK

:CALHist

calibration history

<no>

no = 1 .. 12

R

*STAT:CALH 2

02.03.09 10:42

ACCUMULATION.0.31

:CALMode

mode of the calibration

R

<no>

R

PROOF

R

EXTERNAL

R

NO CAL

*STAT:CALM?

NO CAL

:ERRor

read error

R

*STAT:ERR?

NO ERROR/WARNING

:ERRorHist

<no>

read error history (no =1 .. 12)

R

*STAT:ERRH 1? 20.09.09 09:20:29 W81

:TRIGger

<no>

read settings of trigger (no = 1 .. 2)

Possible values = ON, OFF, GROSSLEAK1,
GROSSLEAK2, NO VALUE, DISABLED
GROSSLEAK1 means leak rate is >5x highest Trigger,
GROSSLEAK2 means leak rate is >100x highest Trig.

R

:VALve

read settings of the valves, 0 = closed; 1 = open

010=STANDBY, 110=READY,
011=FINE, 101=GROSS, 100=CONTAMINATED

R

*STAT:VAL?

VALVE 110

:MEAS

actual status of the measurement
(status not listed here is displayed as a number, see
section

4.2.1

, command 44 „Get Status“.)

R

<value>

value = INIT, STARTSTANDBY, STANDBY,
CONTAMIN, STARTACC, GROSS1ACC, FINE1,
WAITACC, FINE2, GROSS2ACC, READY,
STARTCAR, GROSSCAR, FINECAR, CALCCAR,
REFCAR, WAITPURGE, PURGE

*STAT:MEAS?

READY

*STOP

stop measurement

S