3 error messages binary mode, Error messages binary mode – INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 32

Advertising
background image

32

RS232 Binary Interface Protocol

jin

s

85

e1

4.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

4.3

Error Messages Binary Mode

No.

Name

Explanation

231

RS232disabled

Control location is not RS232

232

RS232Invalid

Command currently not allowed

(i.e. CAL when running up)

234

RS232PW2disable

Password disabled (for INFICON service only)

235

RS232CmdFailed

Execution of command failed

240

RS232Cmd

Command does not exist

243

RS232Len

Number or length of parameter wrong

244

RS232Para

Parameter not in valid range

252

RS232Start

First byte was not 0x05

253

RS232Checksum

Transmitted and calculated checksum differ

254

RS232Timeout

Time out (Command transmission did not finish within
500 ms)

255

RS232Buffer

Overflow in command receive buffer