INFICON T-Guard Leak Detection Sensor Interface description User Manual
Page 29

RS232 Binary Interface Protocol
29
jin
s
85
e1
4.f
m
In
ter
fa
c
e De
scrip
tion
(1
10
1)
116
Get Times
Measurement times
Byte 0:
0=TGross
1=TFine
2=TMeas
3=TPurge
4=WaitPurge
Time in seconds [float]
117
Set Times
Measurement times
Byte 0:
2=TMeas
3=TPurge
4=WaitPurge
Bytes 1..4:
Seconds [float]
118
Proof
Start and confirm
Proof
Byte 0:
1=Start Proof
2=Confirm Proof
121
Set Date Time
Sets date or time
Byte 0:
0=DATE
1=TIME
Bytes 0..2:
Day, month, two-digit-
year for DATE,
Hours, minutes,
seconds for TIME
122
Get Control Location Gets actual control
location
1=LOCAL
2=RS232
3=LOCAL/RS232
4=PLC
5=LOCAL/PLC
7=ALL
123
Set Control Location
Sets actual control
location
Byte 0:
1=LOCAL
2=RS232
3=LOCAL/RS232
4=PLC
5=LOCAL/PLC
7=ALL
(RS232 cannot
switch off itself)
134
Set Lang
Display language
Byte 0:
0=english
1=german
2=french
3=italian
4=portuguese
5=spanish
6=katakana
135
Get Lang
Display language
Byte 0:
0=english
1=german
2=french
3=italian
4=portuguese
5=spanish
6=katakana
136
Get Date
Date
Bytes 0..2:
Day, month, year
137
Get Time
Time
Bytes 0..2:
Hours, minutes,
seconds
No.
Name
Description
Parameter
Data