INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 27

Advertising
background image

RS232 Binary Interface Protocol

27

jin

s

85

e1

4.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

55

Start Cal

Performs calibration,
no need to confirm
afterwards. Please
check new calibration
factor separately.

56

Get Trigger

Trigger value

Byte 0:
1=Trigger 1
2=Trigger 2
Byte 1:
1=sccm
3=mbar l/s
4=Pa m³/s
5=atm cc/s
6=Torr l/s

Leak Rate [float]

57

Set Trigger

Trigger value

Byte 0:

1=Trigger 1

2=Trigger 2
Byte 1:
1=sccm
3=mbar l/s
4=Pa m³/s
5=atm cc/s
6=Torr l/s
Bytes 2..5: [float]

58

Get Mode

Operation mode

Bit 0..6 :
0=accumulation
1=carrier gas
3=cont. mode
Bit 7 :
0=automatic times

59

Set Mode

Operation mode

Byte 0 :
0=accumulation
1=carrier gas
3=cont. mode

62

Get Error Code

Read which error
occurred

error code number
[uchar], see the
corresponding chapter
"Errors and Warnings"
in the Technical
Handbook.

63

Clear Error

Acknowledge and
clear errors

64

GetMinSinceStart

Time since power on

Bytes 0..3:
time in minutes [ulint]

66

Get Test Leak

Test leak value for
calibration

Byte 0:
0=test leak
1=proof leak
Byte 1:
3=mbar l/s
4=Pa m³/s
6=Torr l/s

Leak Rate [float]