Profibus-DP Protocol

5.1

Preface

In order to use Profibus-DP with T-Guard, you must have ordered a T-Guard in the
Profibus version (Cat. No. 540-002). It is not possible to add the Profibus option to
Cat. No. 540-001.

When setting up the Profibus communication you need to use the GSD file provided
by INFICON.

5.2

Process Data Mapping for Cyclic Data Transfer

5.2.1

Read Process Data (PLC-> Leak Detector)

This data is sent periodically from the programmable logic controller to the leak
detector:

Notice

The Profibus-DP protocol is subject to change.
If you are using this protocol, please ask INFICON for an update.

Byte Bit

Name

Meaning

Similar to PLC

Input

Similar to RS232

ASCII cmd.

0

not used

1

not used

2

CLEAR

Transition 0->1:Clears errors and
warnings

CLEAR

*CLS

3

not used

4

CAL

Transition 0->1: Start calibration

CAL

*CAL:START

5

PROOF

Transition 0->1: Start proof

PROOF

6

STOP

Transition 0->1: Stop

STOP

*STOP

7

START

Transition 0->1: Start

START

*START

0

PURGE_STOP

Transition 0->1 = Stop purging the
measurement line

*PURGE:STOP

1 PURGE_START

Transition 0->1 = Start purging the
measurement line

PURGE

*PURGE:START

2

not used

3

not used

4

STANDBY

Transition 0->1: Switch to
stand-by

STANDBY

*STANDBY

5

GROSS/FINE

Switch between GROSS and
FINE (in Cont. Mode only)

GROSS/FINE

*GROSS ON

6

READY

Transition 0->1: Wake up leak
detector from stand-by mode

READY

*STANDBY:STOP

7

not used

0

1