2 measurement cycle – INFICON T-Guard Leak Detection Sensor Interface description User Manual
Page 21

RS232 ASCII Interface Protocol
21
jin
s
85
e1
2.f
m
In
ter
fa
c
e De
scrip
tion
(1
10
1)
2.4.2.2
Measurement Cycle
Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.
Command sent to
T-Guard
Response received
from HLD5000
Content
*STAT?
MEAS
status of the unit
*STAT:MEAS?
READY
last measurement finished
*CONF:MODE CARGAS
OK
switch to carrier gas mode
*CONF:TRIG2ON OFF
OK
turn off trigger 2
*CONF:TIME:AUT ON
OK
turn on automatic times
*CONF:UNIT:FU SCCM
OK
set flow unit to sccm
*CONF:CF 30000
OK
set carrier gas flow to 30000
*CONF:TRIG1:MBAR*L/S
1E-3
OK
set trigger 1 to 1E-3 mbar*l/s
*CONF:TIME:MEAS?
5.0
read measurement time
*START
OK
start measurement
*STAT:MEAS?
READY
measurement finished
*STAT:ERR?
NO ERROR/WARNING no errors
*READ?
2.30E-4
leak rate