2 measurement cycle – INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 21

Advertising
background image

RS232 ASCII Interface Protocol

21

jin

s

85

e1

2.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

2.4.2.2

Measurement Cycle

Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.

Command sent to

T-Guard

Response received

from HLD5000

Content

*STAT?

MEAS

status of the unit

*STAT:MEAS?

READY

last measurement finished

*CONF:MODE CARGAS

OK

switch to carrier gas mode

*CONF:TRIG2ON OFF

OK

turn off trigger 2

*CONF:TIME:AUT ON

OK

turn on automatic times

*CONF:UNIT:FU SCCM

OK

set flow unit to sccm

*CONF:CF 30000

OK

set carrier gas flow to 30000

*CONF:TRIG1:MBAR*L/S
1E-3

OK

set trigger 1 to 1E-3 mbar*l/s

*CONF:TIME:MEAS?

5.0

read measurement time

*START

OK

start measurement

*STAT:MEAS?

READY

measurement finished

*STAT:ERR?

NO ERROR/WARNING no errors

*READ?

2.30E-4

leak rate