Measurement Cycle
Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.
Command sent to
T-Guard
Response received
from HLD5000
Content
*STAT?
MEAS
status of the unit
*STAT:MEAS?
READY
last measurement finished
*CONF:MODE
ACCUMULATE
OK
switch to accumulation mode
*CONF:TRIG2ON OFF
OK
turn off trigger 2
*CONF:TIME:AUT ON
OK
turn on automatic times
*CONF:UNIT:VU LITER
OK
set volume unit to liters
*CONF:AV 10
OK
set free volume to 10
*CONF:TRIG1:MBAR*L/S
5E-4
OK
set trigger 1 to 5E-4 mbar*l/s
*START
OK
start measurement
*STAT:MEAS?
GROSS1ACC
first gross measurement
*STAT:MEAS?
FINE1
first fine measurement
*STAT:MEAS?
WAITACC
accumulating
*READ?
1.0
leak rate not valid
*STAT:MEAS?
GROSS2ACC
second gross measurement
*STAT:MEAS?
FINE2
second fine measurement
*STAT:MEAS?
READY
measurement finished
*STAT:ERR?
NO ERROR/WARNING no errors
*READ?
2.30E-4
leak rate