2 measurement cycle – INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 19

Advertising
background image

RS232 ASCII Interface Protocol

19

jin

s

85

e1

2.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

2.4.1.2

Measurement Cycle

Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.

Command sent to

T-Guard

Response received

from HLD5000

Content

*STAT?

MEAS

status of the unit

*STAT:MEAS?

READY

last measurement finished

*CONF:MODE
ACCUMULATE

OK

switch to accumulation mode

*CONF:TRIG2ON OFF

OK

turn off trigger 2

*CONF:TIME:AUT ON

OK

turn on automatic times

*CONF:UNIT:VU LITER

OK

set volume unit to liters

*CONF:AV 10

OK

set free volume to 10

*CONF:TRIG1:MBAR*L/S
5E-4

OK

set trigger 1 to 5E-4 mbar*l/s

*START

OK

start measurement

*STAT:MEAS?

GROSS1ACC

first gross measurement

*STAT:MEAS?

FINE1

first fine measurement

*STAT:MEAS?

WAITACC

accumulating

*READ?

1.0

leak rate not valid

*STAT:MEAS?

GROSS2ACC

second gross measurement

*STAT:MEAS?

FINE2

second fine measurement

*STAT:MEAS?

READY

measurement finished

*STAT:ERR?

NO ERROR/WARNING no errors

*READ?

2.30E-4

leak rate