INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 11

Advertising
background image

RS232

ASCII Interface

Proto

c

ol

11

jins85e1 2.fm

Interface Description

(1101)

Command

word1

Command

word2

Command

word3

Parameter

Statement

Read/

Set

Example

Response

*CONFig

:RECorder

read/write proof leak

:LINK1_2

analog output 1:

<para>

para = OFF, P1, MANT, EXP, LR_LIN,

LR_LOG, CUR_LIN, CUR_LOG

S
R

*CONF:REC:LINK1_2 EXP
*CONF:REC:LINK1_2?

OK

EXP

:LINK3_4

analog output 2:

<para>

para = OFF, P1, MANT, EXP, LR_LIN,

LR_LOG, CUR_LIN, CUR_LOG

S
R

*CONF:REC:LINK3_4 MANT
*CONF:REC:LINK3_4?

OK

MANT

:SCALE1_2

ascent for logarithmic analog output 1..2:

<para>

para = 0.5V/DEC, 1V/DEC, 2V/DEC,

2.5V/DEC, 3V/DEC, 5V/DEC, 10V/DEC

S
R

*CONF:REC:SCALE1_2 1V/DEC
*CONF:REC:SCALE1_2?

OK

1V/DEC

:SCALE3_4

ascent for logarithmic analog output 3..4:

<para>

para = 0.5V/DEC, 1V/DEC, 2V/DEC,

2.5V/DEC, 3V/DEC, 5V/DEC, 10V/DEC

S
R

*CONF:REC:SCALE3_4 2V/DEC
*CONF:REC:SCALE3_4?

OK

2V/DEC

:upperEXP1_2

max. value for analog output 1..2

<value>

-1 .. -5 = 10

-1

mbarl/s .. 10

-5

mbarl/s

respectively

10

-7

A .. 10

-11

A

S
R

*CONF:REC:EXP1_2 -4
*CONF:REC:EXP1_2?

OK

-4

:upperEXP3_4

max. value for analog output 3..4

<value>

-1 .. -5 = 10

-1

mbarl/s .. 10

-5

mbarl/s

respectively

10

-7

A .. 10

-11

A

S
R

*CONF:REC:EXP3_4 -1
*CONF:REC:EXP3_4?

OK

-1

:PRESSHigh

read/write upper pressure limit in mbar

<value>

S
R

*CONF:PRESSH 370
*CONF:PRESSH?

OK

370

:PRESSLow

read/write lower pressure limit in mbar

<value>

S
R

*CONF:PRESSL 180
*CONF:PRESSL?

OK

180

:STANDBYDel

setting for standby delay:

<value>

00:30

#

30 s; 01:00

#

1 min; 02:00

#

2 min;

05:00

#

5 min; 10:00

#

10 min;

30:00

#

39 min; 60:00

#

1 h; 0 = off

S
S
R

*CONF:STANDBYD OFF
*CONF:STANDBYD 01:00
*CONF:STANDBYD?

OK
OK

01:00