2 carrier gas mode, 1 performing a calibration, Carrier gas mode – INFICON T-Guard Leak Detection Sensor Interface description User Manual
Page 20

20
RS232 ASCII Interface Protocol
jin
s
85
e1
2.f
m
In
ter
fa
c
e De
scrip
tion
(1
10
1)
2.4.2
Carrier Gas Mode
2.4.2.1
Performing a calibration
Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.
Command sent to
T-Guard
Response received
from HLD5000
Content
*STAT:MEAS?
READY
status of the unit
*CONF:MODE CARGAS
OK
switch to carrier gas mode
*CAL:STAT?
NO CAL RUNNING
no calibration running
*CONF:TRIG2ON OFF
OK
turn off trigger 2
*CONF:TIME:AUT ON
OK
turn on automatic times
*CONF:UNIT:FU SCCM
OK
set flow unit to sccm
*CONF:CF 30000
OK
set carrier gas flow to 30000
*CONF:TRIG1:MBAR*L/S
1E-3
OK
set trigger 1 to 1E-3 mbar*l/s
*CONF:TIME:MEAS?
5.0
read measurement time
*CONF:TLR:MBAR*L/S 8E-4 OK
set test leak rate to 8E-4 mbar*l/s
*CAL:START
OK
start calibration measurement
*CAL:STAT?
CAL RUNNING, WAIT
calibration running
*CAL:STAT?
CAL FINISHED, CONFIRM calibration finished
*STAT:ERR?
NO ERROR/WARNING
no errors
*CAL:FAC:OLD?
1.017
read old calibration factor
*CAL:FAC:NEW?
1.098
read new calibration factor
*CAL:QUIT
OK
confirm calibration