INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 9

Advertising
background image

RS232

ASCII Interface

Proto

c

ol

9

jins85e1 2.fm

Interface Description

(1101)

Command

word1

Command

word2

Command

word3

Parameter

Statement

Read/

Set

Example

Response

*CONFig

:HEPERcent

<value>

Helium percentage 10 .. 100%

S
R

*CONF:HEPER 25
*CONF:HEPER?

OK

25

:LANGuage

english, german, espanol, portuguese,

katakana, italiano et francais

<value>

ENGlish; DEUtsch; ESPanol, PORTuguese,

KATakana, ITAliano, FRAncais

S
R

*CONF:LANG ENG
*CONF:LANG?

OK

ENG

:MODE

changing the operation mode resets the time

settings to automatic

<value>

ACCUMULATE; CARGAS; CONTMODE

S

R

*CONF:MODE CARGAS
*CONF:MODE?

OK

CARGAS

:NAME

name of the current parameter set

(replace space by underscore)

S
R

*CONF:NAME 15_L
*CONF:NAME?

OK

15_L

:PLCINLINK

PLC input link for one pin

:<no.>

pin number = 1 .. 6

R/S

<value>

values for PLC input functions:

R/S

NOt_used

R/S

START INV_START

S

R

*CONF:PLCINLINK:2 START
*CONF:PLCINLINK:2?

OK

START

STOP INV_STOP

R/S

START/STOP INV_START/STOP

R/S

CAL INV_CAL

R/S

CLEAR INV_CLEAR

R/S

READY INV_READY

R/S

STANDBY INV_STANDBY

R/S

PROOF INV_PROOF

R/S

PURGE INV_PURGE

R/S

GROSS INV_GROSS

R/S