1 command list binary mode, Command list binary mode – INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 26

Advertising
background image

26

RS232 Binary Interface Protocol

jin

s

85

e1

4.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

4.2.1

Command List Binary Mode

Tip: Often used commands are bold

No.

Name

Description

Parameter

Data

2

Get vacuum
pressure

Fore vacuum
pressure

Byte 0: unit
0=mbar
1=Pa
2=atm
3=Torr

Pressure [float]

5

Get Device ID

Device type

40=T-Guard

8

Get Version

Software version

Byte 0=main version
Byte 1=sub version

10

Get Valves

Valves status

Unsigned int.
V1 0x0100
(normally closed)
V2 0x0200
(normally open)
V3 0x0400
(normally open)
which means:
0=READY
3=GROSS
4=CONTAMINATION
5=FINE
6=STANDBY

16

Get Contrast

Display contrast

Byte 0

38

Get Min Since Start

Uptime

Bytes 0..3: time in
minutes [ulint]

44

Get Status

Gets status

1=StartStandby
2=Standby
3=Contaminated
4=RunUp
5=StartAccumulation
8=NoPurge
10=AccGross1
20=AccFine1
25=AccWait
30=AccFine2
32=AccGross2
36=WaitPurge
38=Purge
40=Ready
50=CGStart
55=CGGross
65=CGFine
70=CGRef
150=CMStart
155=CMGross
160=CMFine
165=CMStop

52

START

Starts measurement

53

STOP

Stops measurement

54

Get Cal State

Gets calibration state

0=No Cal Running
1=Cal Running