1 command list binary mode, Command list binary mode – INFICON T-Guard Leak Detection Sensor Interface description User Manual
Page 26

26
RS232 Binary Interface Protocol
jin
s
85
e1
4.f
m
In
ter
fa
c
e De
scrip
tion
(1
10
1)
4.2.1
Command List Binary Mode
Tip: Often used commands are bold
No.
Name
Description
Parameter
Data
2
Get vacuum
pressure
Fore vacuum
pressure
Byte 0: unit
0=mbar
1=Pa
2=atm
3=Torr
Pressure [float]
5
Get Device ID
Device type
40=T-Guard
8
Get Version
Software version
Byte 0=main version
Byte 1=sub version
10
Get Valves
Valves status
Unsigned int.
V1 0x0100
(normally closed)
V2 0x0200
(normally open)
V3 0x0400
(normally open)
which means:
0=READY
3=GROSS
4=CONTAMINATION
5=FINE
6=STANDBY
16
Get Contrast
Display contrast
Byte 0
38
Get Min Since Start
Uptime
Bytes 0..3: time in
minutes [ulint]
44
Get Status
Gets status
1=StartStandby
2=Standby
3=Contaminated
4=RunUp
5=StartAccumulation
8=NoPurge
10=AccGross1
20=AccFine1
25=AccWait
30=AccFine2
32=AccGross2
36=WaitPurge
38=Purge
40=Ready
50=CGStart
55=CGGross
65=CGFine
70=CGRef
150=CMStart
155=CMGross
160=CMFine
165=CMStop
52
START
Starts measurement
53
STOP
Stops measurement
54
Get Cal State
Gets calibration state
0=No Cal Running
1=Cal Running