INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 28

Advertising
background image

28

RS232 Binary Interface Protocol

jin

s

85

e1

4.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

67

Set Test Leak

Test leak value for
calibration

Byte 0:
0=test leak
1=proof leak
Byte 1:
3=mbar l/s
4=Pa m³/s
6=Torr l/s
Bytes 4..7:
[float]

68

Get Run Hours Basic Uptime Basic

Bytes 0..3:
Time in 0.1hs
[ulint]

70

Get Serial Number

Serial number of
device

Bytes 0..10:
Number (ASCII;
0x20..0x7F)

72

Get Mode

Operation mode

Byte 0:
1=Run up
2=Measure
6=Standby

74

Get Press Limit Low/
High

Get pressure limits

Byte 0:
0=Low
1=High
Byte 1:
0=mbar
1=Pa
2=atm
3=PSI
4=Torr

Pressure Limit [float]

75

Set Press Limit Low/
High

Set pressure limits

Byte 0:
0=Low
1=High
Byte 1:
0=mbar
1=Pa
2=atm
3=PSI
4=Torr
Byte 2..5: [float]

78

Get Cal Fac

Get calibration factor

Factor [float]

79

Set Cal Fac

Set calibration factor

Factor [float]

90

Get Program Version Software version

Byte 0:
Main Version
Byte 1:
Sub Version
Byte 2:
Debug Version

97

Get Filtered
Ion Current

Current in Amps [float]

98

Get Unfiltered Ion
Current

Current in Amps [float]

99

Get Leak Rate

Leak rate

Byte 0:
3=mbar l/s
4=Pa m³/s
6=Torr l/s

Leak Rate [float]
is 1 if no value available

No.

Name

Description

Parameter

Data

Advertising