1 accumulation mode, 1 performing a calibration, Accumulation mode – INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 18

Advertising
background image

18

RS232 ASCII Interface Protocol

jin

s

85

e1

2.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

2.4.1

Accumulation Mode

2.4.1.1

Performing a calibration

Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.

Command sent to

T-Guard

Response received

from HLD5000

Content

*STAT:MEAS?

READY

status of the unit

*CONF:MODE
ACCUMULATE

OK

switch to accumulation mode

*CAL:STAT?

NO CAL RUNNING

no calibration running

*CONF:TRIG2ON OFF

OK

turn off trigger 2

*CONF:TIME:AUT ON

OK

turn on automatic times

*CONF:UNIT:VU LITER

OK

set volume unit to liters

*CONF:AV 10

OK

set free volume to 10

*CONF:TRIG1:MBAR*L/S
1E-3

OK

set trigger 1 to 1E-3 mbar*l/s

*CONF:TLR:MBAR*L/S 8E-4 OK

set test leak rate to 8E-4 mbar*l/s

*CAL:START

OK

start calibration measurement

*CAL:STAT?

CAL RUNNING, WAIT

calibration running

*CAL:STAT?

CAL FINISHED, CONFIRM calibration finished

*STAT:ERR?

NO ERROR/WARNING

no errors

*CAL:FAC:OLD?

1.017

read old calibration factor

*CAL:FAC:NEW?

1.098

read new calibration factor

*CAL:QUIT

OK

confirm calibration