INFICON T-Guard Leak Detection Sensor Interface description User Manual

Page 30

Advertising
background image

30

RS232 Binary Interface Protocol

jin

s

85

e1

4.f

m

In

ter

fa

c

e De

scrip

tion

(1

10

1)

140

Get Background

Background Info from
Display Unit

0=OK
1=moderate
2=bad

144

GetCalEnable

0 = disabled
1 = enabled

145

SetCalEnable

Byte0: 0 = disable
1 = enable

152

Get Stat

Gets status

1=Trigger Status

Bit0 (1)=Trigger1 has
been exceeded
Bit1 (2)=Trigger2 has
been exceeded
Bit4 (16)=Leak Rate is
>5x highest Trigger
Bit5 (32)=Leak Rate is
>100x highest Trigger

154

Get Trigger2 ON

Read if Trigger 2 is
enabled

0=off
1=on

155

Set Trigger2 ON

Enable Trigger 2

0=off

1=on

156

CopyParaSet

Copies parameter set
from Byte0 to Byte1

Byte0: 68=default
0=current set
1..3=internal
11..35=I·Stick
Byte1: 0=current set
1..3=internal
11..35=I·Stick

160

Get Acc Vol

Get accumulation
volume

Byte 0:
0=liters
1=cu feet
2=cu inches
3=ccm

Value [float]

161

Set Acc Vol

Set accumulation
volume

Byte 0:
0=liters
1=cu feet
2=cu inches
3=ccm
Bytes 1..4:
Value [float]

162

Get Car Flow

Get total carrier gas
flow

Byte 0:
0=sccm
1=liters/s

Value [float]

163

Set Car Flow

Set total carrier gas
flow

Byte 0:
0=sccm
1=liters/s
Bytes 1..4:
Value [float]

174

Get Gross

Read if Gross is active
(in Cont. Mode only)

0=off
1=on

175

Set Gross

Set GROSS mode
(in Cont. Mode only)

Byte 0:
0=off
1=on

No.

Name

Description

Parameter

Data