INFICON T-Guard Leak Detection Sensor Interface description User Manual
Page 30

30
RS232 Binary Interface Protocol
jin
s
85
e1
4.f
m
In
ter
fa
c
e De
scrip
tion
(1
10
1)
140
Get Background
Background Info from
Display Unit
0=OK
1=moderate
2=bad
144
GetCalEnable
0 = disabled
1 = enabled
145
SetCalEnable
Byte0: 0 = disable
1 = enable
152
Get Stat
Gets status
1=Trigger Status
Bit0 (1)=Trigger1 has
been exceeded
Bit1 (2)=Trigger2 has
been exceeded
Bit4 (16)=Leak Rate is
>5x highest Trigger
Bit5 (32)=Leak Rate is
>100x highest Trigger
154
Get Trigger2 ON
Read if Trigger 2 is
enabled
0=off
1=on
155
Set Trigger2 ON
Enable Trigger 2
0=off
1=on
156
CopyParaSet
Copies parameter set
from Byte0 to Byte1
Byte0: 68=default
0=current set
1..3=internal
11..35=I·Stick
Byte1: 0=current set
1..3=internal
11..35=I·Stick
160
Get Acc Vol
Get accumulation
volume
Byte 0:
0=liters
1=cu feet
2=cu inches
3=ccm
Value [float]
161
Set Acc Vol
Set accumulation
volume
Byte 0:
0=liters
1=cu feet
2=cu inches
3=ccm
Bytes 1..4:
Value [float]
162
Get Car Flow
Get total carrier gas
flow
Byte 0:
0=sccm
1=liters/s
Value [float]
163
Set Car Flow
Set total carrier gas
flow
Byte 0:
0=sccm
1=liters/s
Bytes 1..4:
Value [float]
174
Get Gross
Read if Gross is active
(in Cont. Mode only)
0=off
1=on
175
Set Gross
Set GROSS mode
(in Cont. Mode only)
Byte 0:
0=off
1=on
No.
Name
Description
Parameter
Data