4 - 7

IP

N 07

4-

28

9L

Composer Operating Manual

4.3.4.2.2 Bulk Flow Control Configuration

Refer to

section 3.1.13 on page 3-11

for Bulk Flow Control specifics.

4.3.5 Description and Purpose of Displays

The following display types are available:

Š

Standard Display Description, see section 4.3.5.1 on page 4-7

Š

Optional - Composer LCD Display Description, see section 4.3.5.2 on page
4-9

4.3.5.1 Standard Display Description

There are nine high-intensity status LEDs on the front panel of the Controller.
See

Figure 4-2

.

NOTE: All LEDs are illuminated for approximately two seconds after the power

switch is cycled on to allow operational verification.

Figure 4-2 Composer front panel

CONCENTRATION . . . . . . . . . . . . A green LED that is illuminated when the

Instrument System is actively controlling
and modifying the voltage on the
CONTROL OUT connector and the
measured concentration is within the
Concentration Window parameter’s value.
The default value of concentration
tolerance is 0.01%

TEMPERATURE . . . . . . . . . . . . . . A green LED that is steadily illuminated

whenever the Transducer’s temperature is
within +0.1

°C to -0.2 °C of the set point. It

flashes slowly when the temperature is
below the range and rapidly when above.

POWER

RECEIVING

SENDING

HALT

WARNING

USER SET ZERO

TEMPERATURE

STEADY

CONCENTRATION

STATUS

COMPOSER

GAS COMPOSITION CONTROLLER

Advertising