Composer Operating Manual

Codes for S and R 99
0 Frequency

Float

1 Amplitude

Float

2 Temperature

Float

3 Concentration

Float

4 BubblerFlow

Float

5 DilutionFlow

Float

6 R2

Float

7 Q

Float

8 Step

Float

9 Error

Float

10 TargetConcentration

Float

From input if configured that way,
otherwise the parameter value

11 Analog Input

Float

Instantaneous reading at last
measurement

12 Digital Input

1 Byte

bit pattern

13 Digital Output

1 Byte

bit pattern

14 Configuration Switches

2 Byte

bit pattern

15 Keyboard Lock

1 Byte

TRUE or FALSE

16 State Status

1 Byte

Same pattern as the current command

17 Drive Level

1 Byte

18 Amp Range

1 Byte

19 Configuration Option

1 Byte

No Option, Bulk Flow, extra output

20 Weight Fraction

Float

Only available if in Bulk Flow

21 MFC Readback

Float

Only available if in Bulk Flow

22 ActualBulkFlow

Float

Only available if in Bulk Flow

23 TargetBubblerFlow

Float

Only available if in Bulk Flow

24 TargetBulkFlow

Float

Only available if in Bulk Flow

25 OptionalAnalogInput

Float

Only available if in Bulk Flow

26 InAmpTrack

byte

1 = Amp Track mode
0 = not in AmpTrack mode

27 ControlLoopType

byte

0 = PID
1 = Fuzzy
2 = PI
3 = I
4 = Open Loop

NOTES:
S99 will not be updated when in sweep.
This has been tested for all codes except the ones that require the option board.
There is a maximum of 30 commands allowed in an R 99.
Formats of all values match the formats in the current "S" commands.
Addition codes can be added as needed.