Composer Operating Manual
6.2.2 Category 2; Acoustic Generator Related
NOTE: PRT stand for Platinum Resistance Thermometer, C stands for
concentration,
µP stands for microphone.
6.2.3 Category 3; Acoustic Receiver Related
Table 6-2 Category 2; Acoustic Generator Related
Failure
Symptom Visibility
Effect on
Accuracy
Drive
µP fails/not
connected
Low amplitude, Noisy C signal
Noise on C
signal
Both
µPs fail/not
connected
Control and Warning LEDs blink,
HALT LED illuminated
N/A
Unbalanced
µP
assembly
Low amplitude or Noisy C signal
Noise on C
signal
Table 6-3 Category 3; Acoustic Receiver Related
Failure
Symptom Visibility
Effect on
Accuracy
Receiver
µP fails/not
connected
Control and Warning LEDs blink,
Error relay opens
N/A
Receiver
µP loses
sensitivity
Noisy C signal at low P
Noise on C
signal
Receiver
µP loses bias No signal, Error relay opens
N/A
Receiver
µP has
unstable bias
Noisy C signal
Noise on C