2 category 2; acoustic generator related, 3 category 3; acoustic receiver related – INFICON Composer Gas Concentration Controller User Manual

Page 135

Advertising
background image

6 - 3

IP

N 07

4-

28

9L

Composer Operating Manual

6.2.2 Category 2; Acoustic Generator Related

NOTE: PRT stand for Platinum Resistance Thermometer, C stands for

concentration,

µP stands for microphone.

6.2.3 Category 3; Acoustic Receiver Related

Table 6-2 Category 2; Acoustic Generator Related

Failure

Symptom Visibility

Effect on
Accuracy

Drive

µP fails/not

connected

Low amplitude, Noisy C signal

Noise on C
signal

Both

µPs fail/not

connected

Control and Warning LEDs blink,
HALT LED illuminated

N/A

Unbalanced

µP

assembly

Low amplitude or Noisy C signal

Noise on C
signal

Table 6-3 Category 3; Acoustic Receiver Related

Failure

Symptom Visibility

Effect on
Accuracy

Receiver

µP fails/not

connected

Control and Warning LEDs blink,
Error relay opens

N/A

Receiver

µP loses

sensitivity

Noisy C signal at low P

Noise on C
signal

Receiver

µP loses bias No signal, Error relay opens

N/A

Receiver

µP has

unstable bias

Noisy C signal

Noise on C
signal

Advertising