5 category 5; user related – INFICON Composer Gas Concentration Controller User Manual
Page 137
6 - 5
IP
N 07
4-
28
9L
Composer Operating Manual
6.2.5 Category 5; User Related
Table 6-5 Category 5; User Related
Failure
Symptom Visibility
Effect On Accuracy
Wrong M and/or
γ
entered
None
Relative to error. Refer
to
M and
γ
are identical for
carrier and precursor
Warning illuminates
No results given.
Reference Zero set at
inappropriate time
None
Relative to delta F
2
from
true zero. Maximum
discrepancy restricted by
limits associated with
designated carrier gas.
Reference Zero left at
factory default
Zero LED off
Minimal
Cell's end caps loosened Facility toxic gas alarms Various
Change of process
pressure without re-zero
None
See dF/dP charts, refer
to
Large barometric
pressure change without
re-zero
None
Up to 0.5 Hz
Change of cell's
temperature without
re-zero
Zero LED off if T
0
is
changed
Proportional to delta T