5 category 5; user related – INFICON Composer Gas Concentration Controller User Manual

Page 137

Advertising
background image

6 - 5

IP

N 07

4-

28

9L

Composer Operating Manual

6.2.5 Category 5; User Related

Table 6-5 Category 5; User Related

Failure

Symptom Visibility

Effect On Accuracy

Wrong M and/or

γ

entered

None

Relative to error. Refer
to

section 3.1.4 on page

3-2

.

M and

γ

are identical for

carrier and precursor

Warning illuminates

No results given.

Reference Zero set at
inappropriate time

None

Relative to delta F

2

from

true zero. Maximum
discrepancy restricted by
limits associated with
designated carrier gas.

Reference Zero left at
factory default

Zero LED off

Minimal

Cell's end caps loosened Facility toxic gas alarms Various

Change of process
pressure without re-zero

None

See dF/dP charts, refer
to

section 3.1.6 on page

3-5

.

Large barometric
pressure change without
re-zero

None

Up to 0.5 Hz

Change of cell's
temperature without
re-zero

Zero LED off if T

0

is

changed

Proportional to delta T