6 category 6; miscellaneous, 3 how to revert to factory default values, 4 if you can not resolve your problem – INFICON Composer Gas Concentration Controller User Manual
Page 138
Advertising
6 - 6
IP
N 07
4-
28
9L
Composer Operating Manual
6.2.6 Category 6; Miscellaneous
6.3 How To Revert To Factory Default Values
Refer to
How To Revert To Factory Default Settings
, located on
.
6.4 If You Can Not Resolve Your Problem
Refer to
section 1.4, How To Contact Customer Support, on page 1-4
Table 6-6 Category 6; Miscellaneous
Failure
Symptom Visibility
Effect on
Accuracy
Small Diaphragm leak
Facility toxic gas alarms
Probably minor
System pressure out of
specified range
a) Over-signal clipping and
Error relay opens
b) Under - Noisy C Signal
c) >1 atmosphere gauge -
subsequent Noisy C signal
@ or near atmospheric
pressures.
a) N/A
b) Noise on C
signal
c) loss of exact
zero reference
Loss of power (Controller
only)
All relays open, C output = 0 N/A
Advertising