6 category 6; miscellaneous, 3 how to revert to factory default values, 4 if you can not resolve your problem – INFICON Composer Gas Concentration Controller User Manual

Page 138

Advertising
background image

6 - 6

IP

N 07

4-

28

9L

Composer Operating Manual

6.2.6 Category 6; Miscellaneous

6.3 How To Revert To Factory Default Values

Refer to

section 4.2.4

,

How To Revert To Factory Default Settings

, located on

page

see page 4-4

.

6.4 If You Can Not Resolve Your Problem

Refer to

section 1.4, How To Contact Customer Support, on page 1-4

.

Table 6-6 Category 6; Miscellaneous

Failure

Symptom Visibility

Effect on
Accuracy

Small Diaphragm leak

Facility toxic gas alarms

Probably minor

System pressure out of
specified range

a) Over-signal clipping and
Error relay opens
b) Under - Noisy C Signal
c) >1 atmosphere gauge -
subsequent Noisy C signal
@ or near atmospheric
pressures.

a) N/A
b) Noise on C
signal
c) loss of exact
zero reference

Loss of power (Controller
only)

All relays open, C output = 0 N/A

Advertising